Technical Information Metrohm NIRS Analyzer Pro Typical Performance The Metrohm NIRS Analyzer Pro is one of the fastest spectrometers on the market with an acquisition time below 1 s. The NIRS PRO is designed to perform in chal enging industrial environments thanks to its rugged enclosure. Indi-cated below is the typical performance of the NIRS Pro as measured with standard materials placed in the analyzer and at sample interface. Parameter Wavelength Range (nm) 1 100–1 650 Detector Indium Gal ium Arsenide (InGaAs) Diode Array Spectral dispersion (nm/pixel) 1.1 5–50 ms / integration time; depends upon sample. Analysis Time Typical result time: 1–15 seconds Wavelength Accuracy (nm) 0.5 Wavelength Precision (nm) < 0.02 Wavelength Stability (nm/°C) < 0.01 Bandwidth (nm) 9.5 Noise (μAu) < 60 Absorbance 0–1.5 Lifetime of the light source (h) Dual lamp system MTBF = 17,500 0.4 Grms at 10–150 Hz in accordance with IEC 60068-2-64 Random vibrations (Grms) 0.4 Grms at 10–1 250 Hz in accordance with international Metrohm standard (additional information available on request) NIRS Pro Process Analyzer Installation Information Metrohm NIRS Analyzer Pro NIRS Pro analyzers are designed to provide precise and accurate results while operating in harsh environments. Please observe the fol owing guidelines when selecting and instal ing your NIRS Pro instrument. Site Requirements Operating Temperature Range -5–40 °C (23–104 °F) with Purge -5–65 °C (23–149 °F) Cooling air flow rate minimum 5 L/min Purge flow rate > 99.9% water free > 99.9% free of oil and fine particles down to 0.3 μm Relative Humidity 10–90% RH, non-condensing IP 69K Ingress Protection according to IEC 60529 and DIN 40050 part 9, NT ELEC 023 Enclosure certification ATEX 20 ready Enclosure material 1.5 mm (lid 2.5 mm) stainless steel EN 1.4301 (SS2333) Weight 25 kg (55 lbs) Dimensions (w × h × d) 420 × 420 × 130 mm (17” × 17” × 5”) Network Connections Local Area Network (LAN) – Ethernet, RJ-45 connection Process Communication Options Digital Protocols MODBUS (TCP/IP or Serial), PROFIBUS VISION I/O Module 12 Analog Outputs, 12 Digital Inputs, 12 Digital Outputs OPC USB or paral el port dongle Electrical Requirements Power Supply Isolated or conditioned supply line Operating Voltage 100–240 VAC Current 2 A Power Consumption 150 W V. www.metrohm.com ohm Applikon B. Subject to modifications Designed by Alyson Lanciki, Metr 8.928.5006EN – 2018-08