Technical Information Metrohm NIRS XDS Process Analyzers: Microbundle Single Point and Multiplexer Typical Performance The XDS is the next generation of Near-Infrared (NIR) spectroscopic instrumentation and is designed to ensure effi-cient and cost-effective NIR method development and implementation. Indicated below is the typical performance of the NIRS XDS Process Analyzer as measured with standard materials placed in the analyzer sample interface. Parameter Measurement Mode Reference Reflectance / Immersion / Transmission Sample Interface Optic probe Optic probe or Flow cel Wavelength Range (nm) 800–2 200 800–2 200 Extended-range Indium Extended-range Indium Detector Gallium Arsenide Gallium Arsenide Data Acquisition Rate 0.5 sec./scan 0.5 sec./scan Spectral Data Interval (nm) 0.5 0.5 Wavelength Accuracy (nm) N/A < 0.08 Wavelength Precision (nm): N/A < 0.004 single instrument Wavelength Precision (nm): N/A < 0.08 instrument to instrument Stray Light < 0.1% < 0.1% Photometric Linearity < 1% of reading < 1% of reading Bandpass 9.0 + 1.0 9.5 + 1.0 Absolute linear range (au) 0–6 0–6 Noise (μau) Measurement Mode Reference Reflectance Immersion Transmission 800–1 100 nm < 40 < 100 < 100 < 40 1 100–1 700 nm < 15 < 30 < 20 < 20 1 700–2 200 nm < 20 < 30 < 20 < 20 NIRS XDS Process Analyzer Installation Information Near-Infrared Spectroscopy (NIRS) XDS analyzers are designed to provide precise and accurate results while operat-ing in harsh environments. Please observe the fol owing guidelines when selecting and instal ing your NIRS XDS instrument. Site Requirements Operating Temperature Range 0–30 °C (32–86 °F) with Vortex cooler (or Purge Control & Vortex cooler) 0–40 °C (32–104 °F) with Air Conditioning (or Purge Control & Air Conditioning) 0–51 °C (0–124 °F) Relative Humidity 10–90% RH, non-condensing NEMA Rating 4X Ingress Protection IP 65 Weight 41.9 kg (92.3 lbs) with Vortex cooler 48.0 kg (105.8 lbs) with Purge Control and Vortex cooler 59.4 kg (130.9 lbs) with Air Conditioning 50.5 kg (111.3 lbs) with Purge Control and Air Conditioning 61.9 kg (136.4 lbs) Dimensions (w × h × d) 485 × 646 × 358 mm (19.09” × 25.43” × 14.09”) with Purge System, fiber hookup area, & other options 879 × 702 × 358 mm (34.60” × 27.64” × 14.09”) Process Communication Options Digital Protocols MODBUS, PROFIBUS, etc. VISION I/O Module Available as 8, 16, or 24 Analog Channels, 24 Digital Channels OPC USB or paral el port dongle Electrical Requirements Power Supply Isolated or conditioned supply line Operating Voltage 120 / 240 V AC switchable Current 2 A / 1.5 A (selectable) with Air Conditioning 5 A / 2.7 A Power Consumption 100 W V. www.metrohm.com ohm Applikon B. -08 Subject to modifications Designed by Alyson Lanciki, Metr 8.928.5004EN – 2017