Suppression in anion chromatography More sensitive analysis of anions and organic acids What is suppression? 02 Suppression plays a key role in the analysis of anions and And now: STREAM – the green way of suppression organic acids using ion-exchange chromatography and Suppression with the MSM is optimized and simplified conductivity detection. even further thanks to STREAM (Suppressor Treatment Reusing Eluent After Measurement). STREAM refers to a Suppression in anion chromatography process whereby, rather than simply disposing of the • decreases the background conductivity of the eluent sup pressed eluent after the detection, the eluate is used • minimizes baseline noise to rinse the regenerated suppressor unit. STREAM also leads • transforms analytes in free anions with protons as to a significant reduction in regene rant consumption. counterions (which involves a remarkable increase in the conductivity signal) The advantages of STREAM at a glance • optimizes the signal-to-noise ratio • No need for ultrapure water as a rinsing solution • increases the detection sensitivity of the measurement • Less waste system • Regenerant consumption reduced by a factor of 3 • Significant reduction in demand for consumables The Metrohm Suppressor Module The patented Metrohm Suppressor Module (MSM) is a As a result, the system can be operated for much longer suppression solution that is just as smart as it is reliable. without requiring any intervention by the user. The Metrohm anion suppressor consists of a small rotor, containing three cartridges filled with cation exchanger resin. While one cart ridge is used for suppression, a re ge-ne ration step is performed on the second one. The third cartridge is rinsed automatically during this time. Thus, a freshly regenerated suppressor cartridge is always available for every new sample. Connection to 1st cartridge Connection to 2nd cartridge Connection to 3rd cartridge 03 Highlights of the Metrohm Suppressor Module for anion chromatography • 10 year warranty (anion suppressor) • 100% pressure stability • 100% resistance to organic solvents • Short conditioning time • Minimum noise of < 0.2 nS/cm • STREAM – the green way of suppression • Cost-effective and robust • Application can be changed at any time with no negative impact on system performance • Excellent signal-to-noise ratio enables anions and organic acids to be analyzed down to the ultratrace range 04 Chemical suppression for anions Sequential suppression for anions The MSM consists of three cartridges filled with cation Sequential suppression combines chemical suppression exchanger material. The first cartridge is used for sup-and CO suppression. For this, a Metrohm CO Suppressor 2 2 pres sion. Simultaneously, the second cartridge is regene- (MCS) is used in addition to the MSM. rated with diluted acid (e.g., sulfuric acid) and the third cartridge is rinsed with the eluate or water. Before each Sequential suppression reduces the background conduc-analysis, the suppressor is rotated 120° so that a freshly tivity further (< 1.5 µS/cm), minimizes the impact of car-regenerated and rinsed cartridge is always available for bonate in the sample matrix (> 98% of carbonate re mo-suppression. ved) and improves the separation of the injection peak from early eluting anions, such as fluoride. This also fur-Reactions that take place during chemical ther increases the sensitivity of the analysis of anions and suppression organic acids. Analyte: The MCS consists of a gas-permeable membrane, which is surrounded by a chamber and contains a vacu um. The Na+Cl– + RSO –H+ g H+Cl– + RSO –Na+ carbonic acid produced during chemical sup pres sion 3 3 breaks down into carbon dioxide and water. The carbon R = substrate dioxide diffuses through the gas-permeable mem brane and is removed. This allows the chemical equi librium to This example of an anion analysis includes a sodium shift more and more towards carbon dio xide and water. analyte counterion. This ion is replaced with a proton As less and less carbonic acid and there fore less disso ci-with an equi valent conductivity that is five times higher. a ted carbonic acid is present, background conductivity is This signifi cantly increases the conductivity of the sample reduced even further. This in turn improves the signal-to-solution and therefore also the signal strength. noise ratio and al ows lower detection limits than those that would be possible with chemical sup pression alone. Salts from weakly dissociated acids (e.g., sodium carbonate/sodium hydrogen carbonate) are used as eluent. The reaction that takes place in the Metrohm CO Suppressor (MCS) 2 Eluent: H+ + HCO – H CO H O + CO (h) 3 2 3 2 2 Na+HCO – + RSO –H+ g H CO + RSO –Na+ 3 3 2 3 3 R = substrate The eluent counterions are also replaced with protons. The carbonic acid that is produced in this way is unstable and only weakly dissociated, meaning that lower background conductivity is measured. Depending on the eluent composition, background conductivity values of 10 to 20 µS/cm are typical for chemical suppression. 05 Precision with every analysis A fresh suppressor cartridge is used for each measure-is delivering its full performance: today, tomorrow, and ment with the MSM. This means that users can be sure for many years to come. they are always working with a cation an exchanger that The chromatograms below demonstrate the outstanding reproducibility of measurements provided by the MSM – here with sequential suppression for anion analysis: 1 2 3 4 5 7 6 30 injections, separation of the standard anions on Metrosep A Supp 5 - 100/4.0, eluent 3.2 mmol/L Na CO / 1.0 mmol/L NaHCO , 2 3 3 flow rate 0.7 mL/min, loop 20 μL, 45 °C, sequential suppression Peak 1 2 3 4 5 6 7 Anions Fluoride Chloride Nitrite Bromide Nitrate Phosphate Sulfate Concentration in mg/L 0.25 0.25 0.25 0.25 0.25 0.25 0.25 Relative standard deviation in % (n = 30) 0.26 0.06 0.42 0.17 0.30 0.46 0.25 Eluate MCS Conductivity detector MSM Peristaltic pump 06 Flow diagram for STREAM with peristaltic pump: the eluate is suppressed in the MSM and MCS before it is detected in the conductivity detector. The regenerant is supplied by a peristaltic pump. Regeneration – two versions Two options are available when it comes to regenerat-ing the suppressor: the new STREAM process and the conventional version. STREAM is the standard version that comes with every Metrohm IC. As described, STREAM reduces the consumption of chemicals and increases convenience for the user. Conventional suppression is used if the eluent is coup led to another detector following conductivity detection (e.g., PCR-UV/VIS or amperometric detection but also IC-MS or IC-ICP/MS couplings). Thus, a second peristaltic channel is necessary to transfer the rinsing water through the suppressor. Free choice of suppressor The suppressor rotor can be freely selected with ion chromatographs from the 940 Professional IC Vario and 930 Compact IC Flex series. Metrohm offers three che mical anion suppressors with different capacities (MSM, MSM-HC, MSM-LC). • The MSM is the suppressor rotor for universal use in routine applications • The MSM-HC is the high-capacity suppressor rotor for gradient applications or isocratic applications with very strong eluents • The MSM-LC is the low-capacity suppressor rotor with a minimum dead volume for applications with 2-mm columns All three suppressor rotors – MSM-HC, MSM, and MSM-LC – have the same design and vary solely in terms of their capacity. Eluate MCS Conductivity detector MSM 800 Dosino 07 Flow diagram for STREAM with Dosino Regeneration: the eluate is suppressed in the MSM and MCS before it is detected in the conductivity detector. An 800 Dosino supplies the regenerant in this system. Regenerant supply For STREAM, the regenerant can be supplied either by a peristaltic pump or an 800 Dosino. The latter of the two versions is known as «Dosino Regeneration» and offers the following advantages: • The 800 Dosino requires less maintenance than the peristaltic pump • MagIC Net can monitor the volume of regenerant consumed and informs the user if the regenerant needs to be refilled • As the peristaltic pump is not required, it can be used for sample preparation techniques such as Inline Ultrafiltration and Inline Dialysis Suppression from a different perspective Until now, suppression has been considered in the con-text of determining anions and organic acids using ion-exchange chromatography and conductivity detection. However, the use of an anion suppressor is recommen-ded if anions are to be determined using UV/VIS detection (in particular between 190–230 nm) too. The sup - pression of the background signal decreases the noise in the UV/VIS detector. In addition, any interferences caused by car bonate in the sample matrix can be avoid-ed; for example, when determining iodide in mi neral water. Lower de tec tion limits can be attained when the suppressor is com bined with the 944 Pro fes sional UV/ VIS Detector Vario. 4.0 An additional application for the suppressor is inverse 3.6 suppression. This is used in the determination of or ga nic 3.2 acids with ion-exclusion chromatography and con duc-2.8 tivity detection, for example. Producing lithium salts from 2.4 with MSM the organic acids in the anion suppressor can signi fi-2.0 cantly increase sensitivity (by a factor of 2-25 de pend-Absorbance [mAU] 1.6 without MSM ing on the pK value of the acid). As a result, this type 1.2 a 0.8 of ion chromatography can even be used to determine 0.4 low concentrations. 0 1 2 3 4 5 6 7 8 9 10 11 12 Time [min] UV/VIS detection at 218 nm with and without suppression: determination of nitrite, bromide, and nitrate Technical information on the anion suppressor rotors MSM MSM-HC MSM-LC Design Robust and chemically resistant micro packed-bed suppressor Approx. 90 min* Approx. 350 min* Approx. 45 min** Capacity (0.45 meq) (1.8 meq) (0.13 meq) Dead volume < 50 µL < 250 µL < 10 µL Regeneration Chemical regeneration with STREAM Maximum flow rate No irreversible damage possible Backpressure No irreversible damage possible, no specific backpressure limitations required for operation Solvent stability 100% solvent stability Temperature range No limit High-capacity suppressor Low-capacity suppressor for gradient applications or with minimum dead Application Universal use isocratic applications with volume for applications very strong eluents with 2-mm columns Factory warranty 10 years *... Standard conditions for Metrosep A Supp 5 - 100/4.0: eluent 3.2 mmol/L Na CO / 1.0 mmol/L NaHCO , flow rate 0.7 mL/min 2 3 3 **... Standard conditions for Metrosep A Supp 10 - 100/2.0: eluent 5.0 mmol/L Na CO / 5.0 mmol/L NaHCO , flow rate 0.2 mL/min 2 3 3 Ordering information Suppressor rotors 6.2832.000 MSM Rotor A 6.2842.000 MSM-HC Rotor A 6.2844.000 MSM-LC Rotor A Accessories 6.2835.010 Connecting piece for MSM-HC and SPM 6.2842.020 Adapter for Suppressor Vario (always required if 6.2832.000 or 6.2844.000 is used in combination with 940 Professional IC Vario or 930 Compact IC Flex) Dosino Regeneration 2.800.0010 800 Dosino 6.5330.190 IC equipment: Dosino Regeneration www.metrohm.com 8 Subject to modifications Layout by Ecknauer+Schoch ASW, printed in Switzerland at Metrohm AG, CH-9100 Herisau 8.000.5134EN – 2015-0