Discover Multi The new Autolab/M204 Multi Autolab/M204 The Multi Autolab/M204 is the newest addition to the SPECIFICATIONS Multi Autolab/M204 Autolab series instruments. The new Multi Autolab/M204 system is based on a user- • Electrode connections 2, 3 and 4 confi gurable multichannel cabinet which can accomodate • Potential range +/- 10 V up to 12 M204 potentiostat/galvanostat boards. • Compliance voltage +/- 20 V The M204 is a 20 V/400 mA potentiostat/galvanostat, • Maximum current +/- 400 mA fi tted with a built-in analog integrator and with analog • Current ranges 100 mA to 10 nA, and digital input/outputs which can be used to interface in 8 decades to external instrumentation. • Potential accuracy +/- 0.2 % • Potential resolution 3 µV Furthermore, the Multi Autolab/M204 instrument can be • Current accuracy +/- 0.2 % complemented by adding extra modules to extend the • Current resolution 0.0003 % functionality of individual M204 boards. This provides the (of current range) means to customize each channel for a specifi c electro- • Input impedance > 100 GOhm chemical measurement. • Potentiostat bandwidth 1 MHz • Computer interface USB Optional modules • Control software NOVA • FRA32M – Electrochemical impedance spectroscopy • pX1000 – pH and temperature measurements • MUX – Multiplexing module for complete cells or individual working electrodes • BA – Dual mode bipotentiostat • EQCM – Electrochemical quartz crystal microbalance Dedicated to research www.metrohm-autolab.com Discover The new Autolab RRDE Autolab RRDE The Autolab Rotating Ring Disc Electrode (RRDE) is the SPECIFICATIONS Autolab RRDE latest addition to the Metrohm Autolab accessories for electrochemical measurements. • Speed control Manual or by NOVA The Autolab RRDE can be used to perform electrochemical software measurements under controlled hydrodynamic conditions. • Motor speed range 100 – 10,000 RPM +/- 1 RPM The Autolab RRDE uses 2 identical friction-less mercury • Acceleration/deceleration 4,000 RPM/s contacts for ultra-low noise measurements and can be • Maximum disc current 500 mA used with any Autolab potentiostat/galvanostat fi tted with • Maximum ring current 50 mA the BA, dual-mode bipotentiostat module. • Contact Double sealed Hg pool • Electrode tips Pt disc/Pt ring, The Autolab RRDE is provided with a motor controller that Au disc/Pt ring, can be operated manually or by the NOVA software. The GC disc/Pt ring rotation rate can be specifi ed between 100 RPM and • Collection effi ciency 24.9 % 10,000 RPM in 1 RPM steps. • Electrode gap 375 µm The Autolab RRDE electrode tips are fi tted with a 5 mm electrode disc and a thin concentric platinum ring. The electrode disc is availalble in Gold, Platinum or Glassy Carbon. The RRDE electrode tips are dimensioned with G, CH-9100 Herisau an electrode gap of 375 µm and a theoretical collection ohm A effi ciency of 24.9 %. witzerland by Metr , printed in S SW Dedicated to research www.metrohm-autolab.com 11 Subject to change Layout by Ecknauer+Schoch A 8.000.5129EN – 2014-