MM500 Moisture analyser for corrosive gases Accurate and reliable, on-line trace moisture analyser for Chlorine, Hydrogen Chloride and other corrosive gases in safe or hazardous area locations This sample system works in conjunction with our range of MM500 analysers for moisture measurement in corrosive gases. Features & Benefits Autoranging from 0.01ppm to 1000ppm No calibration required RS232/485 outputs Inert gas purge ppmv, Dewpoint °C and °F units Fault alarm Sample system IP65/Nema 4 Corrosion resistant materials throughout Designed specifically for corrosive gases, These analysers provide an absolute the MM500 Sample System allows safe, measurement, with accurate dependable accurate and reliable analysis of moisture results over a wide range from 0.01ppm to in corrosive gas mixtures. 1000ppm. Unit of measurement can be selected from ppmv or dew point (°C or °F). The MM500 Moisture Analysers represent the latest advance in moisture Sample wetted parts are made from analysis – designed to be customised to corrosion resistant materials throughout. your application giving precise All non-wetted brackets and materials are measurements while providing a simple, stainless steel. yet affordable analyser. Simply select the instrument configuration and sensor location and let the analyser do All instruments in the range incorporate the rest. microprocessor control, with our high quality electrolytic moisture sensor, to bring flexibility to your process environment. MM500 Moisture Analysers MM513 Bench/Panel Mount 190H x 237W x 410D (mm) 7.9kg MM523 IP66/NEMA 4X Wall Mount/Weatherproof 404H x 328W x 180D (mm) 13.1kg MM533 Rack Mount 4U - 19 inch Houses 1 or 2 analysers MM500 Sample System 178H x 484W x 410D (mm) IP65/Nema 4 9.7kg (single unit) Technical Specifications Measurement Ranges Autoranging from 0.01ppm to 1000ppm and equivalent in Dew Point Other ranges are available with limited accuracy, please consult factory Accuracy ±5% of reading or 0.4 ppm(v) Response Time 90% within 60 seconds Selectable Units ppm(v) / Dew Point °C / Dew Point °F Display Type 5 digit high visibility LED Operating Conditions Sample and ambient temperature: 0–40°C (32–104°F) Sample Connections 1/4 in. OD Maximum Sample Pressure 0.7 – 8.3 Barg Sample Flow 100 cc/min -controlled Power Requirements 115/230 VAC, 50/60 Hz Acceptable Gases Chlorine/HCI and other corrosive gases Options Analogue Outputs Scaleable 0 - 10V, 0 - 100mV and 4 - 20mA or 0 - 20mA all isolated High / Low Alarms 2 voltage free with changeover contacts rated 240V 3A 19” Rack Mount Can be combined with many of our other products in a 19” rack mount configuration Systech Illinois have over 30 years experience of providing analysis solutions for a wide range of industries. From our manufacturing plants in the UK and U.S. we produce gas analysers for industrial process industries, headspace analysers for monitoring gas flushing of food products, and our range of permeation analysers. Systech Instruments Ltd (UK) Illinois Instruments, Inc (U.S) Illinois Instruments (Thailand) Systech Illinois (China) 17 Thame Park Business Centre, 2401 Hiller Ridge Road 6th fl Nopnarong Bldg No7 Room 1105 Forte Building, Wenman Road, Johnsburg, Illinois 60051 Ladprao23, Jatujak, Bangkok 10900 No. 910 Quyang Rd, Hongkou district, Thame, Oxfordshire OX9 3XA U.S.A Thailand Shanghai, China 200434 Tel: +44 (0)1844 216838 Tel: +1 815 344 6212 Tel: +66 (0)2938 0798 Tel: +86 21 65533022 Fax: +44 (0)1844 217220 Fax: +1 815 344 6332 Fax: +66 (0)2938 1058 Fax: +86 21 65539651 E-mail: sales.uk@systechillinois.com E-mail: sales.usa@systechillinois.com E-mail: sales.ap@systechillinois.com Email: info@systechillinois.cn www.systechillinois.com www.systechillinois.com www.systechillinois.com www.systechillinois.cn Systech Illinois reserve the right to change specifications without notice. 05/2014