TECHNICAL SPECIFICATIONS DESCRIPTION Fully automated, direct read, bench top, Discrete analyzer ASSAY TYPE End Point, Kinetic, Differential, Bi chromatic assays 2 racks provided to contain up to 16 reagents as follow: ON-BOARD REAGENTS - 1 rack for 10 reagent position - 1 rack for 6 reagent positions and 10 stock standards - 2 dedicated positions for diluents and other (spike solutions, QC samples…) SAMPLES LOADING 5 racks for continuous loading of samples (126 positions) URGENT SAMPLE ANALYSIS Allowed at any time THROUGHPUT Up to200 tests/hour READING SYSTEM Direct photometry MINIMUM REACTION VOLUME 300 μl Photometer multi - wavelenght optics, double beam Wavelength 12 narrow band interferential filters (9 on board) Source halogen lamp 6V/10W OPTICS SYSTEM Optical Path 10 mm SMARTCHEM 200 Linear range 0.001 - 2.500 Abs Resolution 0.0001 Abs STATE-OF-THE-ART DIRECT READ ed. One mechanic arm provides all sampling operation with : eserv DISCRETE ANALYZER FOR - Capacitive liquid level sensing - Reagent pre-warming at 37°C AUTOMATED CHEMISTRIES SAMPLING SYSTEM - Automatic probe washing - Sample Volume range : 3.0 - 300 μl (1μl incr.) - Reagent Volume : 3.0 - 300 μl (1μl incr.) - Sample pre-dilution - Sample rerun with or without dilution DILUTOR Automatic dilution in pre and post-dilution Circular 60 semi-disposable cuvettes tray: y be changed without prior notice. All rights r REACTION PLATE - Automatic cuvette washing - Continuous monitoring of cuvette quality - Incubation temperature: 37°C DIMENSIONS 42 x 65 x 100 cm (H x W x L) es. Specifications ma WEIGHT 65 Kg actual pictur OPTIONAL ACCESSORIES - Bar code reader - Nitrate module - ISE detector - Peltier chiller for reagent module oup. Non contr Digestion system: Easy digest Distillation system: Easy Dist Ready to use reagents S ALLIANCE Gr COMPLEMENTARY PRODUCTS 15 AM © April 20 AMS France 10 avenue Charles de Gaulle 95740 Frepillon FRANCE NEW ORDER ONLINE Tel. : (+33) 134 18 71 10 reagents and consumables. Fax : (+33) 139 60 72 39 www.amsalliance.com http://webshop.amsalliance.com plaquette-smartchem200-UK-A4.indd 1 10/04/2015 09:15 SMARTCHEM 200 Micro pipettor-dilutor Washing station Smartwash FULLY AUTOMATED, DIRECT READ, THE ADVANTAGES DISCRETE ANALYZER TO MAXIMIZE OF DIRECT READ PRODUCTIVITY FOR ROUTINE AND DISCRETE ANALYSIS SPECIAL CHEMISTRIES, UP TO 200 TESTS/HOUR • Each reaction takes place in its own optical cuvette • True sample blanking performed directly in each cuvette • No flow cells • No baseline stabilization Cadmium module for • No manual manifold changeover Nitrate analysis and/or • No excessive reagent consumption • Automatic reagent blanking to evaluate reagent quality • Pre and Post dilution of samples and calibration standards • Automatic spike preparation • Automatic method changeover • Analyze samples in “batch” mode or “random” mode ISE module for pH, conductivity and On-board reagents Long-life cuvettes with integrated The Smartchem 200 combines the latest technology in redox analysis Cuvette Quality Assurance System discrete analysis with an innovative, unique design that guarantees accurate and reproducible results. The Smartchem 200 is the first discrete analyzer to utilize DIRECT READ : FILTER WHEEL USER FRIENDLY the Cuvette Quality Assurance System, performing an MORE SIMPLICITY & SECURITY • 12 optical filters (9 on board), plus 1 dark filter, ranging from integrated QC check of each cuvette prior to use. In addition SMARTSTART 340nm to 880 nm to automated analysis the Smartchem 200 offers automated The Smartchem 200 utilizes direct read photometry utilizing a true discrete, direct read measurement • Others filters optionally available system Quality Control. SOFTWARE system and reusable, high purity synthetic optical The Smartchem 200 is also the first discrete analyzer to cuvette system eliminating compromised sample utilize a patented, dedicated sample prep module for the COMPUTER CONTROLLED MICRO PIPETTOR measurement. determination of NOx by Cadmium reduction. Designed for The Smartchem 200 operating software runs under Windows • Accurate and precise dispensing of all liquids multi-tasking and high throughput rates, the NOx module SMARTWASH : ELIMINATE CARRYOVER XP, Windows Vista or Windows 7. Its bright graphical user • Probe cleaned after every liquid contact - prevents cross also provides automated cadmium regeneration. interface simplifies routine work and offers the operator a DURING SAMPLING contamination and carryover concise guide through the operating functions. Now, the instrument provides also a pH, Conductivity and • Microliter reagent and sample consumption Redox module. • Level sensor makes probe washing easy, with a NEW SOFTWARE MAIN FEATURES • Liquid level sensor provides warnings for low reagent and standard 2mm immersion depth sample levels • Potential contaminants go the way they need to • Automatic start-up flow – out of the sample probe! • Automatic shutdown INTERACTIVE GRAPHICAL “DASHBOARD” • Three stage process: clean the inside of the probe, NITRATE/NITRITE OR ISE MODULE • Barcode for reagents PROVIDES A USER-FRIENDLY INTERFACE: clean the washing well, then clean the outside of • Automated cadmium coil regeneration for Nitrate & Nitrite • Barcode for samples the probe analysis, complies with US EPA Method # 353.2 and SM • Addition of new sample or method to a sample during the run SELECT METHOD(S) AND TYPE IN SAMPLE IDS • Optical integrity of each cuvette quantitatively 4500-NO3-F verified before each use • Reanalyse key • ISE (Ion Selective Electrodes) module for pH, conductivity and FOLLOW THE DASHBOARD AND PLACE • Provides an infinite number of reaction vessels for • Accurate estimation of completion of analysis redox analysis THE SAMPLES, REAGENTS, AND STOCK analysis • Calibration points recalculation • You can choose between each module or both STANDARD IN THE AUTOSAMPLER • Economical & environmentally friendly: long lasting • Last accepted calibration overlay START ANALYSIS WITH A SIMPLE CLICK, ALL cuvettes, minimize hazardous waste and solid • Pre-dilution wastes PHOTOMETER ANALYTICAL PROCEDURES FROM WORKING • Standard curves • High resolution, dual beam digital detector • Ready to read 2D barcoding in reagents kits STANDARD DILUTION TO SAMPLE ANALYSIS, SAMPLE LOADING • Expanded absorbance range to 2.500 ABS • Computer controlled micro pipettor CUVETTE WASHING AND SYSTEM QC , ARE • 5 universal, separate racks for continuous loading • 5 detection modes: endpoint, fixed time, differential, • NELAC - EPA compliant FULLY AUTOMATED AND DO NOT NEED of samples, calibrators and controls bichromatic, and kinetic • Real-time presentation of sample and reagent status ANY OPERATOR ACTION • Up to 200 tests/hour including ISE • Dynamic monitoring of test execution plaquette-smartchem200-UK-A4.indd 2 10/04/2015 09:16