Treated Water Specifications MODEL PURELAB flex 1 PURELAB flex 1 PURELAB flex 2 PURELAB flex 3 PURELAB flex 4 With Purification Pack Daily volume >10 liters >10 liters >10 liters <10 liters >10 liters Delivery flow rate – maximum Up to 2 l/min Up to 2 l/min Up to 2 l/min Up to 2 l/min Up to 2 l/min Reverse osmosis make up flow rate N/A N/A N/A Up to 10 l/hour N/A Inorganics (resistivity @25°C) As per feedwater 18.2 MΩ-cm 18.2 MΩ-cm 18.2 MΩ-cm 18.2 MΩ-cm Organics (TOC) – typical Dependant on Dependant on <5ppb 1 <5ppb 1 <5ppb 1 feedwater feedwater Direct from internal reservoir N/A N/A N/A Type III / RO water 2 Type III / RO water 2 Bacteria – typical (when fitted with POU Filter) <1 CFU/10ml <1 CFU/10ml <1 CFU/10ml <1 CFU/10ml <1 CFU/10ml Bacteria – typical (when fitted with Biofilter) <1 CFU/10ml <1 CFU/10ml <1 CFU/10ml <1 CFU/10ml <1 CFU/10ml Endotoxin (when fitted with Biofilter) N/A <0.001 EU/ml <0.001 EU/ml <0.001 EU/ml <0.001 EU/ml DNase (when fitted with Biofilter) N/A N/A <20 pg/ml <20 pg/ml <20 pg/ml RNase (when fitted with Biofilter) N/A N/A <0.002 ng/ml <0.002 ng/ml <0.002 ng/ml Dimensions and weights Dimensions Width 236mm, Depth 374mm, Height minimum 900mm, Width 236mm, Depth 470mm, Height Height maximum 1020mm minimum 900mm, Height maximum 1020mm Operational weight 10kg (22lb) 10.5kg (23.1lb) 11kg (24.2lb) 23kg (57.3lb) 23kg (57.3lb) Installation Bench/wall Bench/wall Bench/wall Bench/wall Bench/wall Feedwater requirement Source Originally from potable supply, then pre treated. Potable Tap Water Originally from potable Preferably reverse osmosis (RO) or filtered service supply, then pre treated. deionization (SDI) or distilled. Preferably reverse osmosis (RO) or filtered service deionization (SDI) or distilled. Conductivity <1 µS/cm 3 <1 µS/cm 3 <1 µS/cm 3 <2000 µS/cm 4 <30 µS/cm Contaminant Hardness N/A N/A N/A <350 ppm as CaCO TBC 3 Free Chlorine <0.05 ppm Cl <0.05 ppm Cl <0.05 ppm Cl <0.5 ppm Cl <0.05 ppm Cl 2 2 2 2 2 Chloramine <0.05 ppm Cl <0.05 ppm Cl <0.05 ppm Cl <0.2 ppm Cl <0.05 ppm Cl 2 2 2 2 2 Total Chlorine <0.05 ppm Cl <0.05 ppm Cl <0.05 ppm Cl <0.5 ppm Cl <0.05 ppm Cl 2 2 2 2 2 Silica <2 ppm <2 ppm <2 ppm <30 ppm SiO <2 ppm SiO 2 2 Carbon Dioxide Maximum <0.1 ppm <0.1 ppm <0.1 ppm <30 ppm <30 ppm (recommended <20 ppm) (recommended <20 ppm) Fouling index <1 <1 <1 <10 <1 Iron / Manganese N/A N/A N/A <0.05 ppm N/A Organics (TOC) N/A N/A N/A <2 ppm <50 ppb recommended Particulates A 0.2 micron membrane pre filter is recommended for all non-RO feeds to extend N/A point-of-use filter life Temperature 4 - 40°C (Recommended 10 - 25°C) Pure and ultrapure laboratory Flowrate (requirement at 15°C) >2 l/min >2 l/min >2 l/min Up to 75 l/hr Up to 60 l/hr Drain requirements (gravity fall with air gap) None required None required None required >90L/hr >70L/hr water from a single system 3 If feedwater is in the range of 1 µS/cm – 30 µS/cm you will need a pre-conditioning cartridge 4 Purification pack life may vary with feedwaters >1400 µS/cm Feedwater pressure Maximum 22psi (1.5bar) 22psi (1.5bar) 22psi (1.5bar) 90psi (6bar) 5 90psi (6bar) Minimum Flooded suction Flooded suction Flooded suction 30psi (2bar) 5 1psi (0.0 7bar) 5 If <60psi (4 bar) a separate boost pump is recommended Electrical requirements Mains Input 100-240V ac, 50-60Hz System control voltage (not including 24V dc pumps and UV) Power consumption during peak demand 120VA 120VA 120VA 100VA 100VA Noise level <40dBa Your laboratory water specialists ELGA has been a trusted name in water purification for Water is the most common reagent used in today’s over 50 years, pioneering innovative technologies and laboratories for research and testing applications and award winning product design for our customers. Part we understand how important accurate consistent of the world’s leading water services company Veolia test results are. The award winning PURELAB flex range Water Solutions and Technologies, ELGA have the most is designed around your needs, delivering accuracy, comprehensive range of water purification systems flexibility and ease of use with an innovative and available today for laboratory research, healthcare and ergonomic design. diagnostics. Find a PURELAB flex to match your application PURELAB flex 2 PURELAB flex 3&4 PURELAB flex 1 PURELAB flex 3&4 PURELAB flex 1 Daily volume >10 Liters <10 Liters >10 Liters <10 Liters >10 Liters Water quality Type I ultrapure water Type II/Type II+ pure water Type III RO water Feedwater source Pre-purified Pre-purified Pre-purified Pre-purified Pre-purified water or tap water water or tap water water Typical Electrochemistry AAS (Atomic Absorption Autoclave feed applications Electrophoresis Spectrophotometry) Feed to ultrapure water systems GFAAS (Graphite Furnace Atomic Buffer and media preparation Hydroponics Absorption Spectrophotometry) Electrophysiology Plant growth cabinets HPLC FAAS (Flame Atomic Absorption Stability chambers IC (Ion Chromatography) Spectrophotometry) Steam generators ICPAES (Inductively Coupled Plasma Feed to ultrapure water systems Atomic Emission Spectrometry) Glassware washing/rinsing Sterilizer feed ICPMS (Inductively Coupled Plasma Mass General chemistry Spectrometry) Histology Mammalian and bacterial cell culture Microbiological analysis Molecular biology RIA (Radioimmunoassay) / ELISA Plant tissue culture (Enzyme Linked Immunoabsorbant Qualitative analyses Assay) Water purification made easy Intuitive to use, ergonomic handset design Clear water purity display for absolute confidence as you dispense Flexible dispensing in four different ways: Realtime TOC monitoring for critical • Variable flow – drop by drop or up to 2 liters per minute analytical applications • Autovolume dispense from 50ml to 60 liters and repeat dispensing Prioritized information • Hands free with optional foot pedal shown at all times (system • Locked dispense for glassware filling status, TOC, alarm) POU filters for multiple applications • Remove endotoxins, DNase, RNase and bacteria • Conforms to international water standards e.g. CLSI, CLRW, ISO 3696: Grade 1,2,3, ASTM D1193-06, Pharmacopeia USP, EP and JP Optional accessories • Wall mounting bracket • Foot pedal • Thermal printer • Point of Use filters • Leak sensor Designed for today’s laboratories Multiple dispense positioning, wall, bench, height adjustable Customized settings to arm, hand-held dispensing suit your application Designed for simple easy installation 185nm/254nm UV lamp for low TOC water production needed for organic sensitive application Purification pack quick and easy to replace Easy-access for routine maintenance Data capture via USB for system performance validation Quick and easy sanitization Minimize microbial growth to give you absolute confidence in the water purity. The simple procedure involves one single sanitization pack so no harsh chemicals need to be ordered, Global network, local support handled or mixed. No additional consumables need to be replaced during this short process. Our highly trained service engineers will apply their expertise to the installation, validation and maintenance of your water purification system. We aim to provide a first-time-fix service. Inside the PURELAB flex PURELAB flex 2 Process Flow Diagram Water Purity Sensor Water Purity and 185/254 nm Purification Temperature Dispense UV Lamp Pack Sensor Handset Feedwater 0.2µm Point Inlet of Use Filter Water Purity and (Pre-purified or Biofilter Temperature Water) (optional) Sensor Recirculation Pump PURELAB flex 3 Process Flow Diagram Water Purity Reverse Sensor Osmosis CVF Composite Water Vent Filter Purification Purity and 185/254 nm Pack Temperature Dispense Reservoir UV Lamp Sensor Handset 0.2µm Point of Use Filter Water Purity and or Biofilter Temperature (optional) Sensor Feedwater Inlet Manual Recirculation (potable tap water) Outlet Pump Want to know more? Find out more or book your demonstration at www.purelabflex.com Commitment to quality Assured safety PURELAB flex systems are supplied with a Certificate The PURELAB flex is compliant with CE directives of Conformity, ensuring they have been manufactured associated with safety and electromagnetic and tested at ELGA LabWater Global Operations, whose compatibility and is approved and tested by an Quality Management System is approved by Lloyds accredited external company. The PURELAB flex Register Quality Assurance (LRQA) and complies with is also compliant with the requirements of BS EN ISO 9001:2008. IEC/EN61010-1 and UL61010-1 and is marked ELGA’s manufacturing facility has been approved by LRQA with an ETL label showing compliance. and complies with the environmental management system ISO 14001:2004. The PURELAB flex has been designed to conform to the requirements of the Waste Electrical & Electronic Equipment (WEEE) Directive and the Restriction of Hazardous Substance (RoHS) Directive. Treated Water Specifications MODEL PURELAB flex 1 PURELAB flex 1 PURELAB flex 2 PURELAB flex 3 PURELAB flex 4 With Purification Pack Daily volume >10 liters >10 liters >10 liters <10 liters >10 liters Delivery flow rate – maximum Up to 2 l/min Up to 2 l/min Up to 2 l/min Up to 2 l/min Up to 2 l/min Reverse osmosis make up flow rate N/A N/A N/A Up to 10 l/hour N/A Inorganics (resistivity @25°C) As per feedwater 18.2 MΩ-cm 18.2 MΩ-cm 18.2 MΩ-cm 18.2 MΩ-cm Organics (TOC) – typical Dependant on Dependant on <5ppb 1 <5ppb 1 <5ppb 1 Water Purity feedwater feedwater Sensor Direct from internal reservoir N/A N/A N/A Type III / RO water 2 Type III / RO water 2 Water Bacteria – typical (when fitted with POU Filter) <1 CFU/10ml <1 CFU/10ml <1 CFU/10ml <1 CFU/10ml <1 CFU/10ml Purity and 185/254 nm Purification Bacteria – typical (when fitted with Biofilter) <1 CFU/10ml <1 CFU/10ml <1 CFU/10ml <1 CFU/10ml <1 CFU/10ml Temperature Dispense UV Lamp Pack Sensor Handset Endotoxin (when fitted with Biofilter) N/A <0.001 EU/ml <0.001 EU/ml <0.001 EU/ml <0.001 EU/ml DNase (when fitted with Biofilter) N/A N/A <20 pg/ml <20 pg/ml <20 pg/ml Feedwater 0.2µm Point RNase (when fitted with Biofilter) N/A N/A <0.002 ng/ml <0.002 ng/ml <0.002 ng/ml Inlet of Use Filter 1 Water Purity and Dependant on feedwater 2 (Equal to or >) Type III RO Water (Pre-purified or Biofilter Temperature Dimensions and weights Water) (optional) Sensor Dimensions Width 236mm, Depth 374mm, Height minimum 900mm, Width 236mm, Depth 470mm, Height Height maximum 1020mm minimum 900mm, Height maximum 1020mm Recirculation Operational weight 10kg (22lb) 10.5kg (23.1lb) 11kg (24.2lb) 23kg (57.3lb) 23kg (57.3lb) Pump Installation Bench/wall Bench/wall Bench/wall Bench/wall Bench/wall Feedwater requirement Source Originally from potable supply, then pre treated. Potable Tap Water Originally from potable Preferably reverse osmosis (RO) or filtered service supply, then pre treated. deionization (SDI) or distilled. Preferably reverse osmosis (RO) or filtered service deionization (SDI) or distilled. Conductivity <1 µS/cm 3 <1 µS/cm 3 <1 µS/cm 3 <2000 µS/cm 4 <30 µS/cm Contaminant Hardness N/A N/A N/A <350 ppm as CaCO TBC 3 Free Chlorine <0.05 ppm Cl <0.05 ppm Cl <0.05 ppm Cl <0.5 ppm Cl <0.05 ppm Cl 2 2 2 2 2 Chloramine <0.05 ppm Cl <0.05 ppm Cl <0.05 ppm Cl <0.2 ppm Cl <0.05 ppm Cl 2 2 2 2 2 Total Chlorine <0.05 ppm Cl <0.05 ppm Cl <0.05 ppm Cl <0.5 ppm Cl <0.05 ppm Cl 2 2 2 2 2 Silica <2 ppm <2 ppm <2 ppm <30 ppm SiO <2 ppm SiO 2 2 Carbon Dioxide Maximum <0.1 ppm <0.1 ppm <0.1 ppm <30 ppm <30 ppm (recommended <20 ppm) (recommended <20 ppm) Fouling index <1 <1 <1 <10 <1 Iron / Manganese N/A N/A N/A <0.05 ppm N/A Organics (TOC) N/A N/A N/A <2 ppm <50 ppb recommended Particulates A 0.2 micron membrane pre filter is recommended for all non-RO feeds to extend N/A point-of-use filter life Temperature 4 - 40°C (Recommended 10 - 25°C) Flowrate (requirement at 15°C) >2 l/min >2 l/min >2 l/min Up to 75 l/hr Up to 60 l/hr Drain requirements (gravity fall with air gap) None required None required None required >90L/hr >70L/hr 3 If feedwater is in the range of 1 µS/cm – 30 µS/cm you will need a pre-conditioning cartridge 4 Purification pack life may vary with feedwaters >1400 µS/cm Feedwater pressure Maximum 22psi (1.5bar) 22psi (1.5bar) 22psi (1.5bar) 90psi (6bar) 5 90psi (6bar) Minimum Flooded suction Flooded suction Flooded suction 30psi (2bar) 5 1psi (0.0 7bar) 5 If <60psi (4 bar) a separate boost pump is recommended Electrical requirements Mains Input 100-240V ac, 50-60Hz System control voltage (not including 24V dc pumps and UV) Power consumption during peak demand 120VA 120VA 120VA 100VA 100VA Noise level <40dBa ELGA LabWater tel: +44 (0) 203 567 7300 • fax: +44 (0) 203 567 7205 • info@elgalabwater.com • www.elgalabwater.com ELGA is the global laboratory water brand name of Veolia. The information contained in this document is the property of VWS (UK) Ltd, trading as ELGA LabWater, and is supplied without liability for errors or omissions. © VWS (UK) Ltd. 2015 – All rights reserved. ELGA® and PURELAB® are registered trademarks of VWS (UK) Ltd LITR38975-06